Pulse Programmers

Devices

Device Tested Library
Pulse Blaster ESR 500 Pro 06/2021 SpinCore SpinAPI
Insys FM214x3GDA (as multichannel TTL pulse generator) 03/2025 Atomize_ITC libs
Pulse Blaster Micran 12/2023

The device is available via ctypes. The original C library was written by SpinCore Technologies.

The Insys device is available via ctypes. The original library can be found here.

Functions

pulser_name()

pulser_name()    # -> str; device name

This function returns device name.


pulser_pulse(**kargs)

# Default DETECTION pulse, no phase cycling
pulser_pulse()

# A named microwave pulse at 100 ns, length 100 ns
pulser_pulse(
    name='P0', channel='MW',
    start='100 ns', length='100 ns',
    delta_start='0 ns', length_increment='0 ns')

The keyword arguments:

Keyword Default Description
name 'P0' name of the pulse
channel 'DETECTION' channel string (CH0, CH1, …, CH20)
start '0 ns' start time of the pulse (ns, us, ms, s)
length '100 ns' length of the pulse (ns, us, ms, s)
delta_start '0 ns' start time increment of the pulse (ns, us, ms, s)
length_increment '0 ns' pulse length increment (ns, us, ms, s)
phase_list [] phase cycling sequence (+x, -x, +y, -y)

This function sets a pulse with specified parameters. The default argument is name = 'P0', channel = 'DETECTION', start = '0 ns', length = '100 ns', delta_start = '0 ns', length_increment = '0 ns', phase_list = []. The pulse sequence will be checked for overlap. In the auto defence mode (default option; can be changed in the config file) channels AMP_ON and LNA_PROTECT will be added automatically according to the delays indicated in the config file. In this mode AMP_ON and LNA_PROTECT pulses will be joined in one pulse if the distance between them is less than 12 ns (can be changed in the config file).

Allowed channels: 'DETECTION', 'TRIGGER', 'AMP_ON', 'LNA_PROTECT', 'MW', '-X', '+Y', 'TRIGGER_AWG', 'AWG', 'LASER', 'SYNT2', 'CH10', …, 'CH20'

Allowed phase_list values: '+x', '-x', '+y', '-y'

Output format (time args): 'number' + 'ns' | 'us' | 'ms' | 's'

Range (Pulse Blaster ESR 500 Pro): pulse length 10 ns1900 ns; sequence ≈ 10 s max

Range (Insys FM214x3GDA): pulse length 3.2 ns1900 ns; sequence ≈ 10 s max

In the case of Insys FM214x3GDA start, length, delta_start, and length_increment will be rounded to a multiple of 3.2.

In the case of Pulse Blaster ESR 500 Pro and Pulse Blaster Micran the DETECTION pulse may carry a phase_list. This list declares the acquisition phase cycle and is used by pulser_acquisition_cycle() whenever that function is called without an explicit acq_cycle argument; passing acq_cycle directly still overrides it. Alternatively, the acquisition phases can be indicated directly in pulser_acquisition_cycle() and the DETECTION pulse left without a phase_list. The plain TRIGGER channel must always have an empty phase_list. In the case of Insys FM214x3GDA a phase_list of DETECTION pulse is used to phase cycle the data.


pulser_update()

pulser_update()    # commit pulse-sequence changes to the programmer

This function updates a pulse sequence and sends instructions to the pulse programmer. It has to be called after changes have been applied to pulses either via any of the pulser functions or by changing a pulse property directly. Only by calling the function the changes are committed and the real pulses will change.


pulser_next_phase()

pulser_next_phase()    # advance all pulses to the next phase

This function switches all pulses to the next phase. The phase sequence is declared in the pulser_pulse() in the form of phase_list = ['-y', '+x', '-x', '+x', ...]. By repeatedly calling the function one can run through the complete list of phases for the pulses. The length of all phase lists specified for different MW pulses has to be the same. This function also immediately updates the pulse sequence, as it is done by calling pulser_update(). The first call of the function corresponds to the first phase in the phase_list argument of the pulser_pulse().


pulser_acquisition_cycle(data1, data2, acq_cycle=None)

# Two-step phase cycle (+x, -x)
i, q = pulser_acquisition_cycle(
    np.array([1, 0]), np.array([0, 1]),
    acq_cycle=['+x', '-x'])

# acq_cycle omitted: the phase_list of the DETECTION pulse is used
i, q = pulser_acquisition_cycle(np.array([1, 0]), np.array([0, 1]))

This function can be used to shorten the syntax for acquisition in the case of phase cycling. The arguments are (i) two numpy arrays from a quadrature detector, (ii) array of mathematical operations to perform. Data arrays can be both 2D and 1D, representing, respectively, the case of raw oscillograms or integrated data. The length of acq_cycle array and the 1D arrays or the amount of the individual oscillograms in the 2D array should be equal. The data arrays will be treated inside the function as a complex number:

answer = np.zeros( data1.shape ) + 1j*np.zeros( data2.shape )

The symbol at the index J of the acq_cycle array means that the corresponding values from the data arrays will be added with the following factor to the resulting array:

Allowed operations: '+x', '-x', '+y', '-y'

Symbol Factor Operation
+x +1 answer = answer + data1[J] + 1j*data2[J]
-x -1 answer = answer - data1[J] - 1j*data2[J]
+y +1j answer = answer + 1j*data1[J] - data2[J]
-y -1j answer = answer - 1j*data1[J] + data2[J]

The output of the function is the real and imaginary parts of the answer array after complete cycle of mathematical transformations. These can be both 1D and 2D arrays, depending on the shape of the input data arrays.

For Pulse Blaster ESR 500 Pro and Pulse Blaster Micran the acq_cycle argument is optional. When it is omitted (or an empty list is given) the function falls back to the phase_list declared on the DETECTION pulse; an explicitly supplied acq_cycle always takes precedence.

Although this function is available for Insys FM214x3GDA, it is better to use a modified version of digitizer_get_curve(). In this case acquisition phases should be given directly in the phase list key argument of the DETECTION pulse. For Insys FM214x3GDA the function has a modified signature pulser_acquisition_cycle(data1, data2, points, phases, adc_window, acq_cycle=['+x'], lo=None, hi=None) and is normally called internally by digitizer_get_curve(): the data is taken from the running on-board accumulators (the legacy data1 and data2 arguments are accepted but ignored), and the keywords lo and hi indicate the range of points to recompute. The factor table above applies to the acq_cycle argument in the same way.


pulser_repetition_rate(*r_rate)

pulser_repetition_rate()          # -> str (query)
pulser_repetition_rate('2 Hz')    # set to 2 Hz

This function queries (if called without argument) or sets (if called with one argument) the repetition rate of the pulse sequence. If there is an argument it will be set as a repetition rate. If there is no argument the current repetition rate is returned as a string. The maximum available repetition rate depends on the total length of the pulse sequence.

Max: 5 MHz


pulser_shift(*pulses)

pulser_shift()    # shift all active pulses by their delta_start
pulser_shift('P0', 'P1') # shift only the named pulses

This function can be called with either no argument or with a list of comma separated pulse names (i.e. 'P0', 'P1'). If no argument is given the start time of all pulses that have a nonzero delta_start and are currently active (do not have a length of 0) are shifted by their corresponding delta_start value. If there is one argument or a list of comma separated pulse names only the start time of the listed pulses are changed. Calling this function also resets the phase (if specified in the argument phase_list of the pulser_pulse()) to the first phase in the phase_list.


pulser_increment(*pulses)

pulser_increment()        # increment all active pulses by their length_increment
pulser_increment('P0')    # increment only the named pulses

This function can be called with either no argument or with a list of comma separated pulse names (i.e. 'P0', 'P1'). If no argument is given the lengths of all pulses that have a nonzero length_increment and are currently active (do not have a length of 0) are incremented by their corresponding length_increment value. If there is one argument or a list of comma separated pulse names only the lengths of the listed pulses are changed. Calling this function also resets the phase (if specified in the argument phase_list of the pulser_pulse()) to the first phase in the phase_list.


pulser_redefine_start(*, name, start)

pulser_redefine_start(name='P0', start='100 ns')

# Multiple pulses at once
pulser_redefine_start(name=['P0', 'P1'], start=['0 ns', '320 ns'])

This function should be called with two keyword arguments, namely name and start. The first argument specifies the name of the pulse as a string. The second argument defines a new value of pulse start as a string in the format number + [' ms', ' us', ' ns']. Arguments can be either single strings or lists, for example: name = ['P0', 'P1'], start = ['0 ns', '320 ns']. The main purpose of the function is non-uniform sampling. Please note, that the function does not update the pulse programmer. The pulser_update() function should be called to apply changes.

In the case of Insys FM214x3GDA start will be rounded to a multiple of 3.2.


pulser_redefine_delta_start(*, name, delta_start)

pulser_redefine_delta_start(name='P0', delta_start='10 ns')

# Multiple pulses
pulser_redefine_delta_start(
    name=['P0', 'P1'], delta_start=['0 ns', '32 ns'])

This function should be called with two keyword arguments, namely name and delta_start. The first argument specifies the name of the pulse as a string. The second argument defines a new value of delta_start as a string in the format number + [' ms', ' us', ' ns']. Arguments can be either single strings or lists, for example: name = ['P0', 'P1'], delta_start = ['0 ns', '32 ns']. The main purpose of the function is non-uniform sampling. Please note, that the function does not update the pulse programmer. The pulser_update() function should be called to apply changes.

In the case of Insys FM214x3GDA delta_start will be rounded to a multiple of 3.2.


pulser_redefine_length_increment(*, name, length_increment)

pulser_redefine_length_increment(name='P2', length_increment='2 ns')

# Multiple pulses
pulser_redefine_length_increment(
    name=['P0', 'P1'], length_increment=['0 ns', '3.2 ns'])

This function should be called with two keyword arguments, namely name and length_increment. The first argument specifies the name of the pulse as a string. The second argument defines a new value of length increment as a string in the format number + [' ms', ' us', ' ns']. Arguments can be either single strings or lists, for example: name = ['P0', 'P1'], length_increment = ['0 ns', '3.2 ns']. The main purpose of the function is non-uniform sampling. Please note, that the function does not update the pulse programmer. The pulser_update() function should be called to apply changes.

In the case of Insys FM214x3GDA length_increment will be rounded to a multiple of 3.2.


pulser_reset()

pulser_reset()    # reset all pulses and update the pulse programmer

The function switches the pulse programmer back to the initial state (including phase) in which it was in at the start of the experiment. This function can be called only without arguments. It includes the complete functionality of pulser_pulse_reset(), but also immediately updates the pulse programmer as it is done by calling pulser_update().

Note

This function is not available for Insys FM214x3GDA. The function pulser_pulse_reset() can be used instead.


pulser_pulse_reset(*pulses)

pulser_pulse_reset()        # reset all pulses (incl. phases)
pulser_pulse_reset('P1')    # reset only the named pulses

This function switches the pulse programmer back to the initial state in which it was in at the start of the experiment. This function can be called with either no argument or with a list of comma separated pulse names. If no argument is given all pulses are reset to their initial states (including phases). The function does not update the pulser, if you want to reset all pulses and also update the pulser use the function pulser_reset() instead.


pulser_stop()

pulser_stop()    # stop the pulse programmer

This function stops the pulse programmer. The function should always be called at the end of an experimental script in the case of Pulse Blaster ESR 500 Pro.

Note

This function is not available for Insys FM214x3GDA. The function pulser_close() must be used instead.


pulser_state()

pulser_state()    # -> str; current programmer state

This function queries the pulse programmer state and should be called only without arguments.

Note

This function is only available for Pulse Blaster ESR 500 Pro.


pulser_visualize()

pulser_visualize()    # opens a 2D plot of the current pulse sequence

This function visualizes the pulse sequence as 2D plot and should be called only without arguments.


pulser_pulse_list()

pulser_pulse_list()    # -> list of str; the declared pulse sequence

This function should be called only without arguments and it returns the declared pulse sequence as an array.


pulser_open()

pulser_open()    # open the board for use

This function should be called only without arguments and is only available for Insys FM214x3GDA and Pulse Blaster Micran. In the case of Insys FM214x3GDA, the function should be used after defining pulses and repetition rate with pulser_pulse() and pulser_repetition_rate().


pulser_close()

pulser_close()    # gracefully close the board

This function should be called only without arguments and is only available for Insys FM214x3GDA and Pulse Blaster Micran. The function must be used at the end of an experimental script to gracefully close the board.


pulser_default_synt(num)

pulser_default_synt(1)    # select synthesizer 1
pulser_default_synt(2)    # select synthesizer 2

This function should be called only with one argument and selects the default sources for microwave pulse generation.

Allowed: 1, 2

Note

This function is only available for Insys FM214x3GDA.


pulser_phase_reset()

pulser_phase_reset()    # reset the phase index to zero

This function resets the phase index to zero in order to start phase cycling once again.

Note

This function is only available for Pulse Blaster ESR 500 Pro and Pulse Blaster Micran.


pulser_clear()

pulser_clear()    # clear the device module's pulse array and status flags

This is a special function for clearing pulse array {self.pulse_array} and other status flags of the device module. The function is usually used in GUI applications that use the device module.


pulser_full_stop()

pulser_full_stop()    # zero out all pulse instructions

This is a special function for a complete stop of the pulse programmer, which means all the pulse instructions will be set to zero.

Note

This function is only available for Pulse Blaster Micran.


pulser_test_flag(flag)

pulser_test_flag('None')    # normal mode
pulser_test_flag('test')    # test mode

This is a special function for changing test mode. The function is usually used in GUI applications that use the device module.

Allowed: 'None', 'test'