Arbitrary Waveform Generators

Devices

Device Tested Connection
Spectrum M4I 6631 X8 07/2021 PCIe
Insys FM214x3GDA 03/2025 PCIe

The original library for Spectrum M4I 6631 X8 was written by Spectrum. The path to the library header files (pyspcm.py, spcm_tools.py) and the card device node are read from the device configuration file (the [SPECIFIC] section, keys header_dir and device), so they can be set per machine without editing the module:

[SPECIFIC]
header_dir = /path/to/python/header/of/Spectrum/library
device = /dev/spcm0

The Insys FM214x3GDA device is available via ctypes. The original library can be found here.

Functions

awg_name()

awg_name()    # -> str; device name

This function returns device name.


awg_setup()

awg_setup()    # write all settings into the AWG card

This function writes all the settings modified by other functions to the AWG card. The function should be called only without arguments. One needs to initialize the settings before calling awg_update(). The default settings (if no other function was called) are the following: Sample clock is 1250 MHz; Clock mode is 'Internal'; Reference clock is 100 MHz; Card mode is 'Single'; Trigger channel is 'External'; Trigger mode is 'Positive'; Loop is infinity; Trigger delay is 0; Enabled channels are CH0 and CH1; Amplitude of CH0 is '600 mV'; Amplitude of CH1 is '533 mV'; Number of segments is 1; Card memory size is 64 samples; Buffer is empty.

Note

This function is not available for Insys FM214x3GDA. This is done by pulser_open() function.


awg_update()

awg_update()    # run the AWG card

This function redefines the buffer (in case the function like awg_shift() has been called) and runs the AWG card. The function should be called only without arguments.

Note

This function is not available for Insys FM214x3GDA. This is done by pulser_update() function.


awg_stop()

awg_stop()    # stop the AWG card

This function stops the AWG card and should be called only without arguments. If an infinite number of loops is defined by the awg_loop() function, the awg_stop() should always be called before redefining the buffer by the awg_update(). If a finite number of loops is defined, the card will stop automatically.

Note

This function is not available for Insys FM214x3GDA. There is no need to stop this DAC.


awg_close()

awg_close()    # close the AWG driver

This function closes the AWG driver and should be called only without arguments. The function should always be called at the end of an experimental script, since the AWG card driver is opened during whole experiment in order to achieve high rate of buffer updating. It is STRONGLY recommended to add a graceful closing of the card to the experimental scripts for the case of an abrupt termination of the process. As a possible option, one can use signal library:

import signal
import atomize.device_modules.Spectrum_M4I_6631_X8 as spectrum

awg = spectrum.Spectrum_M4I_6631_X8()

def cleanup():
    awg.awg_stop()
    awg.awg_close()
    sys.exit(0)

signal.singal(signal.SIGTERM, cleanup)

# AWG setup and experimental script
#
#

Note

This function is not available for Insys FM214x3GDA. This is done with pulser_close().


awg_pulse(**kargs)

# 40 ns 200 MHz sine pulse with pi/2 phase
awg_pulse(name='P0', channel='CH0', func='SINE',
          frequency='200 MHz', phase=pi/2, length='40 ns')

This function sets a pulse with specified parameters for 'Single', 'Multi', and 'Single Joined' card mode. The AWG card buffer will be filled according to key arguments of the awg_pulse() function.

Keyword Default Description
name 'P0' Pulse name
channel 'CH0' Channel string
func 'SINE' Function type
frequency '200 MHz' Frequency of the pulse
phase 0 Phase of the pulse (in radians)
phase_list [] Phase cycling sequence
delta_phase 0 Phase increment of the pulse (in radians) for 'Single' and 'Multi' card mode
length '16 ns' Pulse length
sigma '0 ns' Sigma value for GAUSS pulses
length_increment '0 ns' Pulse length and sigma increment
start '0 ns' Pulse start for the 'Single Joined' card mode
delta_start '0 ns' Pulse delta start for the 'Single Joined' card mode
amplitude 100 Additional coefficient for adjusting pulse amplitudes
n 1 Special coefficient for WURST and SECH/TANH pulses determining the steepness of the amplitude function
b 0.02 Parameter for SECH/TANH pulse determining the truncation parameter in 1/ns

A channel should be one of the following ['CH0', 'CH1']. The frequency should be in MHz, the minimum value is 0 MHz, maximum is 280 MHz. For WURST and SECH/TANH pulse the frequency argument should be a tuple ('center_freq MHz', 'sweep_range MHz'), i.e. ('0 MHz', '100 MHz'). The scaling factor for length and sigma key arguments should be one of the following ['ns','us','ms']. The minimum available length and sigma of the pulse is 0 ns. The maximum available length and sigma of the pulse is 1900 ns. For 'SINE', 'BLANK', 'WURST', and 'SECH/TANH' function parameter sigma has no meaning. For 'GAUSS' function parameter sigma is a sigma of Gaussian. For 'SINC' function a combination of parameters length and sigma specifies the width of the SINC pulse, i.e. length = '40 ns' and sigma = '10 ns' means that SINC pulse will be from -4pi to +4pi. Function 'BLANK' is an empty pulse. Function 'WURST' is a wideband, uniform rate, smooth truncation pulse. Function 'SECH/TANH' is a wideband sech/tanh pulse. The length_increment keyword affects both the length and sigma of the pulse. The scaling factor for start and delta_start key arguments should be one of the following ['ns','us','ms']. The amplitudes of the AWG card channels will be set according to the value of the amplitude keyword argument. The amplitude keyword argument should be less than or equal to 100% and should indicate the amplitude as a percentage of the maximum available level. The n parameter determines the steepness of the amplitude function of the WURST and SECH/TANH pulses. For other functions it has no meaning. The b parameter (in 1/ns) determines the truncation parameter of the SECH/TANH pulse. For other functions it has no meaning.

It is recommended to first define all pulses and then define the settings of the AWG card. To write the settings awg_setup() function should be called. To run specified pulses the awg_update() function should be called.

Key argument delta_phase defines a pulse phase shift and has no meaning for 'Single Joined' card mode, since the phase of the pulse will be calculated automatically. Key arguments start and delta_start define a pulse position for 'Single Joined' card mode and has no meaning for 'Single' or 'Multi' card mode.

Note

Pulse settings for Insys FM214x3GDA correspond to 'Single' card mode. In addition for this device arguments 'start', 'length', 'delta_start', and 'length_increment' will be rounded to a multiple of 3.2 and for all AWG pulses there should be a corresponding trigger pulse, generated by pulser_pulse() function.

Allowed channels: 'CH0', 'CH1'

Allowed functions: 'SINE', 'GAUSS', 'SINC', 'BLANK', 'WURST', 'SECH/TANH'

Allowed phase_list values: '+x', '-x', '+y', '-y'

Range frequency: 0 MHz280 MHz

Range length / sigma: 0 ns1900 ns


awg_pulse_sequence(**kargs)

# pulse train of three pulses with 10 kHz repetition rate
# and a third pulse shifting by 40 ns at each of 800 points
awg_pulse_sequence(pulse_type=['SINE', 'GAUSS', 'SINE'],
                   pulse_start=[0, 160, 320], pulse_delta_start=[0, 0, 40],
                   pulse_length=[40, 120, 40], pulse_phase=['+x', '+x', '+x'],
                   pulse_sigma=[40, 20, 40], pulse_frequency=[50, 200, 40],
                   n=[20, 20, 20], b=[0.02, 0.02, 0.02],
                   number_of_points=800, loop=10, rep_rate=10000)

This function sets a pulse sequence with specified parameters for 'Sequence' card mode. The AWG card buffer will be filled according to key arguments of the awg_pulse_sequence() function. There is no default arguments, that is why all keywords must be specified.

Keyword Description
pulse_type List of function types per pulse
pulse_start List of pulse starts in ns
pulse_delta_start List of pulse delta starts in ns
pulse_length List of pulse lengths in ns
pulse_phase List of pulse phases
pulse_sigma List of pulse sigmas in ns
pulse_frequency List of pulse frequencies in MHz
number_of_points Total number of points in a pulse sequence, taking into account shifting of the pulses
loop Number of repetitions of each point
rep_rate Repetition rate of the pulse sequence in Hz
n List of n parameters in arb. u.
b List of b parameters in 1/ns

The minimum available length and sigma of the pulse is 0 ns. The maximum available length and sigma of the pulse is 1900 ns. For WURST and SECH/TANH pulses the frequency should be a tuple ('Center_freq MHz', 'sweep_freq MHz'), i.e. ('0 MHz', '100 MHz'). For 'SINE', 'BLANK', 'WURST', and 'SECH/TANH' functions parameter sigma has no meaning. For 'GAUSS' function parameter sigma is a sigma of Gaussian. For 'SINC' function a combination of parameters length and sigma specifies the width of the SINC pulse in the same manner as the function awg_pulse(). Function 'BLANK' is an empty pulse. Function 'WURST' is a wideband, uniform rate, smooth truncation pulse. Function 'SECH/TANH' is a wideband sech/tanh pulse. The n parameter determines the steepness of the amplitude function of the WURST and SECH/TANH pulse. For other functions it has no meaning. The b parameter (in 1/ns) determines the truncation parameter of the SECH/TANH pulse. For other functions it has no meaning.

Please note, that each new call of the awg_pulse_sequence() function will redefine the pulse sequence. pulse_start array must be sorted. A number of enabled channels should be defined before awg_pulse_sequence().

To write the settings awg_setup() function should be called. To run a specified pulse sequence the awg_update() function should be called. After going through all the buffer the AWG card will be stopped.

Note

This function is not available for Insys FM214x3GDA.

Allowed pulse_type values: 'SINE', 'GAUSS', 'SINC', 'BLANK', 'WURST', 'SECH/TANH'

Allowed pulse_phase values: '+x', '-x', '+y', '-y'


awg_shift(*pulses)

awg_shift()              # shift phase of all active pulses by their delta_phase
awg_shift('P0', 'P1')    # shift only specified pulses

This function can be called with either no argument or with a list of comma separated pulse names (i.e. 'P0', 'P1'). In the 'Single' or 'Multi' card mode, if no argument is given the phase of all pulses that have a nonzero delta_phase and are currently active (do not have a length of 0) are shifted by their corresponding delta_phase value. If there is one argument or a list of comma separated pulse names only the phase of the listed pulses are changed.

In the 'Single Joined' card mode, the start values of the specified pulses are shifted by the defined delta_start values. Phases of the pulses are not changed, since they are calculated from the pulse positions automatically.

For the 'Sequence' card mode, the function has no meaning. Calling this function also resets the phase (if specified in the argument phase_list of the awg_pulse()) to the first phase in the phase_list.


awg_increment(*pulses)

awg_increment()        # increment length & sigma of all active pulses by their increment
awg_increment('P0')    # increment only the pulse named 'P0'

This function can be called with either no argument or with a list of comma separated pulse names (i.e. 'P0', 'P1'). In the 'Single', 'Multi' or 'Single Joined' card mode, if no argument is given the lengths and sigmas of all pulses that have a nonzero increment and are currently active (do not have a length of 0) are incremented by their corresponding increment value. If there is one argument or a list of comma separated pulse names only the lengths and sigmas of the listed pulses are changed.

Please, note that the function always keeps the ratio length/sigma for GAUSS and SINC pulses. For instance, if length = '64 ns', sigma = '16 ns', and increment = '10 ns' after calling awg_increment() once, the parameters will be length = '104 ns', sigma = '26 ns', and increment = '10 ns'.

For the 'Sequence' card mode, the function has no meaning. Calling this function also resets the phase (if specified in the argument phase_list of the awg_pulse()) to the first phase in the phase_list.


awg_next_phase()

awg_next_phase()    # switch all declared pulses to the next phase

This function switches all pulses to the next phase. The phase sequence is declared in the awg_pulse() in the form of phase_list = ['-y', '+x', '-x', '+x', ...]. Argument '+x' means that the phase of the pulse will be equal to phase declared in the awg_pulse(). Argument '-x' corresponds to the phase shifted by pi radians in comparison with the phase declared in the awg_pulse(). Argument '+y' corresponds to pi/2 radians shift. Argument '-y' – 3pi/2 radians. By repeatedly calling the function one can run through the complete list of phases for the pulses. The length of all phase lists specified for different pulses has to be the same. This function also immediately updates the AWG card buffer, as it is done by calling awg_update(). The first call of the function corresponds to the first phase in the phase_list argument of the awg_pulse().


awg_redefine_delta_phase(*, name, delta_phase)

# change delta_phase setting of the 'P0' pulse to pi radians
awg_redefine_delta_phase(name='P0', delta_phase=pi)

This function should be called with two keyword arguments, namely name and delta_phase. The first argument specifies the name of the pulse as a string. The second argument defines a new value of delta_phase in radians. Arguments can be either single strings or lists, for example: name = ['P0', 'P1'], delta_phase = [0, np.pi]. The main purpose of the function is non-uniform sampling. Please note, that the function does not update the AWG card. The awg_update() function should be called to apply changes. The function has no meaning for the 'Single Joined' and 'Sequence' card mode, since phases of the pulses are calculated from the pulse positions automatically.


awg_redefine_phase(*, name, phase)

# change phase setting of the 'P0' pulse to pi radians
awg_redefine_phase(name='P0', phase=pi)

This function should be called with two keyword arguments, namely name and phase. The first argument specifies the name of the pulse as a string. The second argument defines a new value of phase in radians. Arguments can be either single strings or lists, for example: name = ['P0', 'P1'], phase = [0, np.pi]. The main purpose of the function is phase cycling. Please note, that the function does not update the AWG card. The awg_update() function should be called to apply changes. The function has no meaning for the 'Sequence' card mode. One should redefine all the sequence instead.


awg_redefine_amplitude(*, name, amplitude)

# change amplitude setting of the 'P0' pulse to 50%
awg_redefine_amplitude(name='P0', amplitude=50)

This function should be called with two keyword arguments, namely name and amplitude. The first argument specifies the name of the pulse as a string. The second argument defines a new value of the amplitude keyword argument of the awg_pulse() as a percentage of the maximum available level. The amplitude should be in the range of 0–100%. Arguments can be either single strings or lists, for example: name = ['P0', 'P1'], amplitude = [50, 100]. Please note, that the function does not update the AWG card. The awg_update() function should be called to apply changes. The function has no meaning for the 'Sequence' card mode. One should redefine all the sequence instead.


awg_redefine_frequency(*, name, freq)

# change frequency setting of the 'P0' pulse to '10 MHz'
awg_redefine_frequency(name='P0', freq='10 MHz')

This function should be called with two keyword arguments, namely name and frequency. The first argument specifies the name of the pulse as a string. The second argument defines a new value of frequency as a string, i.e. '100 MHz' or a list of string ('0 MHz', '100' MHz) for WURST and SECH/TANH pulses, see awg_pulse() for more details. Arguments can be either single strings or lists, for example: name = ['P0', 'P1'], freq = ['10 MHz', ('0 MHz', '50 MHz')]. Please note, that the function does not update the AWG card. The awg_update() function should be called to apply changes. The function has no meaning for the 'Sequence' card mode. One should redefine all the sequence instead.


awg_redefine_delta_start(*, name, delta_start)

# change delta_start setting of the 'P0' pulse to 10 ns
awg_redefine_delta_start(name='P0', delta_start='10 ns')

This function should be called with two keyword arguments, namely name and delta_start. The first argument specifies the name of the pulse as a string. The second argument defines a new value of delta_start as a string. Arguments can be either single strings or lists, for example: name = ['P0', 'P1'], delta_start = ['0 ns', '32 ns']. The main purpose of the function is non-uniform sampling. Please note, that the function does not update the AWG card. The awg_update() function should be called to apply changes. The function has no meaning for the 'Single', 'Multi', and 'Sequence' card mode, since in these modes the start of the pulse is determined by the trigger event.

Output format: 'number' + 'ns' | 'us' | 'ms'


awg_redefine_length_increment(*, name, length_increment)

# change length increment setting of the 'P2' pulse to 10 ns
awg_redefine_length_increment(name='P2', length_increment='10 ns')

This function should be called with two keyword arguments, namely name and length_increment. The first argument specifies the name of the pulse as a string. The second argument defines a new value of length increment as a string. Arguments can be either single strings or lists, for example: name = ['P0', 'P1'], length_increment = ['0 ns', '32 ns']. The main purpose of the function is non-uniform sampling. Please note, that the function does not update the AWG card. The awg_update() function should be called to apply changes. The function has no meaning for the 'Sequence' card mode. One should redefine all the sequence instead.

Output format: 'number' + 'ns' | 'us' | 'ms'


awg_add_phase(*, name, add_phase)

# add pi radians to the 'P0' pulse
awg_add_phase(name='P0', add_phase=pi)

This function should be called with two keyword arguments, namely name and add_phase. The first argument specifies the name of the pulse as a string. The second argument defines a value of phase in radians to add. Arguments can be either single strings or lists, for example: name = ['P0', 'P1'], add_phase = [0, np.pi]. The main purpose of the function is phase cycling. Please note, that the function does not update the AWG card. The awg_update() function should be called to apply changes. The function has no meaning for the 'Sequence' card mode. One should redefine all the sequence instead.


awg_reset()

awg_reset()    # reset all pulses to initial state and update the AWG card

This function switches the AWG card back to the initial state in which it was in at the start of the experiment. This function can be called only without arguments. It includes the complete functionality of awg_pulse_reset(), but also immediately updates the AWG card as it is done by calling awg_update(). The function has no meaning for the 'Sequence' card mode. One should redefine all the sequence instead.

Note

This function is not available for Insys FM214x3GDA. The function awg_pulse_reset() can be used instead.


awg_pulse_reset(*pulses)

awg_pulse_reset()        # reset all pulses to their initial state
awg_pulse_reset('P1')    # reset only the pulse named 'P1'

This function switches the AWG card back to the initial state in which it was in at the start of the experiment. This function can be called with either no argument or with a list of comma separated pulse names. If no argument is given all pulses are reset to their initial states (including phases). If there is one argument or a list of comma separated pulse names only the listed pulses are returned back to the initial state. The function does not update the AWG card, if you want to reset all pulses and and also update the AWG card use the function awg_reset() instead. The function has no meaning for the 'Sequence' card mode. One should redefine all the sequence instead.


awg_number_of_segments(*segments)

awg_number_of_segments()     # -> int (query)
awg_number_of_segments(2)    # set number of segments to 2

This function queries or sets the number of segments for 'Multi' card mode. In order to set the number of segments higher than 1, the AWG card should be in 'Multi' mode. If there is no argument the function will return the current number of segments. If there is an argument the specified number of segments will be set. Default value is 1.

Note

This function is not available for Insys FM214x3GDA.

Range: 1200


awg_channel(*channel)

awg_channel()                 # -> str (query)
awg_channel('CH0', 'CH1')     # enable output from CH0 and CH1

This function enables output from the specified channel or queries enabled channels. If there is no argument the function will return the currently enabled channels. If there is an argument the output from the specified channel will be enabled. Default option is when both channels are enabled.

Note

This function is not available for Insys FM214x3GDA, only two channels are available.

Allowed: 'CH0', 'CH1'


awg_sample_rate(*s_rate)

awg_sample_rate()        # -> int (query)
awg_sample_rate(1250)    # set sample rate to 1250 MHz

This function queries or sets the AWG card sample rate (in MHz). If there is no argument the function will return the current sample rate. If there is an argument the specified sample rate will be set. Default value is 1250 MHz. Please note that sample rate affects the delay between a trigger event and the AWG card output. The delay is determined in samples and can be found in the documentation.

Note

This function is not available for Insys FM214x3GDA, the default sample rate is 1250 MHz.

Range: 50 MHz1250 MHz


awg_clock_mode(*mode)

awg_clock_mode()              # -> str (query)
awg_clock_mode('Internal')    # set internal clock mode

This function queries or sets the AWG card clock mode. If there is no argument the function will return the current clock mode setting. If there is an argument the specified clock mode will be set. According to the documentation, the internal sampling clock is generated in default mode by a programmable high precision quartz. The external clock input of the M3i/M4i series is fed through a PLL to the clock system. Therefore the input will act as a reference clock input thus allowing to either use a copy of the external clock or to generate any sampling clock within the allowed range from the reference clock. Due to the fact that the driver needs to know the external fed in frequency for an exact calculation of the sampling rate the reference clock should be set by the awg_reference_clock() function. Default setting is 'Internal'.

Note

This function is not available for Insys FM214x3GDA.

Allowed: 'Internal', 'External'


awg_reference_clock(*ref_clock)

awg_reference_clock()       # -> int (query)
awg_reference_clock(100)    # set reference clock to 100 MHz

This function queries or sets the AWG card reference clock in MHz for 'External' mode of the awg_clock_mode() function. If there is no argument the function will return the current reference clock. If there is an argument the specified reference clock will be set. Default value is 100 MHz.

Note

This function is not available for Insys FM214x3GDA.

Range: 10 MHz1000 MHz


awg_card_mode(*mode)

awg_card_mode()           # -> str (query)
awg_card_mode('Multi')    # set 'Multi' card mode

This function queries or sets the AWG card mode. If there is no argument the function will return the current card mode setting. If there is an argument the specified card mode will be set. According to the documentation, in the 'Single' mode a data from on-board memory will be replayed on every detected trigger event. The number of replays can be programmed by loops. 'Single Joined' mode is a modification of the 'Single' mode with a possibility of defining more than one pulse. In this mode, all defined pulses are combined into a sequence of pulses according to their start position, determined by start keyword of the awg_pulse() function. Please note, that if two channels are enabled the pulse sequence for the second channel will be generated automatically using a phase shifting specified in the config file. In the 'Multi' mode every detected trigger event replays one data block (segment). Segmented memory is available only in 'External' trigger mode. In the 'Sequence' mode it is possible to define a whole pulse sequence by the awg_pulse_sequence() function. The pulse sequence has a specified number of points that is looped specified times. Switching between points is achieved using a trigger event. Please note, that if two channels are enabled the pulse sequence for the second channel will be generated automatically using a phase shifting specified in the config file. Default setting is 'Single'.

Note

This function is not available for Insys FM214x3GDA, it operates in the mode close to 'Single'.

Allowed: 'Single', 'Multi', 'Single Joined', 'Sequence'


awg_trigger_channel(*channel)

awg_trigger_channel()              # -> str (query)
awg_trigger_channel('Software')    # set 'Software' trigger

This function queries or sets the AWG card trigger channel. If there is no argument the function will return the current trigger channel. If there is an argument the specified channel will be used as trigger. Trigger channel 'External' corresponds to 'Trg0' channel of the AWG card. Default setting is 'External'.

Note

This function is not available for Insys FM214x3GDA.

Allowed: 'Software', 'External'


awg_trigger_mode(*mode)

awg_trigger_mode()              # -> str (query)
awg_trigger_mode('Positive')    # trigger detection for positive edges

This function queries or sets the AWG card trigger mode. If there is no argument the function will return the current trigger mode. If there is an argument the specified trigger mode will be set. Mode 'Positive' corresponds to trigger detection for positive edges (crossing level 0 from below to above). 'Negative' to trigger detection for negative edges (crossing level 0 from above to below). 'High' to trigger detection for HIGH levels (signal above level 0). 'Low' to trigger detection for LOW levels (signal below level 0). Default setting is 'Positive'.

Note

This function is not available for Insys FM214x3GDA.

Allowed: 'Positive', 'Negative', 'High', 'Low'


awg_loop(*loop)

awg_loop()     # -> int (query)
awg_loop(0)    # set infinite number of loops

This function queries or sets the number of loops for 'Multi' or 'Single' card mode. If there is no argument the function will return the current number of loops. If there is an argument the specified number of loops will be set. A setting 0 means an infinite number of loops. Default value is 0.

Note

This function is not available for Insys FM214x3GDA.

Range: 0100000 (0 = infinite)


awg_trigger_delay(*delay)

awg_trigger_delay()             # -> str (query)
awg_trigger_delay('51.2 ns')    # set trigger delay to 51.2 ns

This function queries or sets the AWG card trigger delay. If there is no argument the function will return the current trigger delay. If there is an argument the specified trigger delay will be set. The delay step is 32 sample clock. If an input is not divisible by 32 sample clock the delay will be rounded and a warning message will be printed. Default value is '0 ns'.

Note

This function is not available for Insys FM214x3GDA.

Output format: 'number' + 'ns' | 'us' | 'ms'


awg_amplitude(*amplitude)

awg_amplitude('CH0')                          # -> str (query for channel)
awg_amplitude('CH0', '600', 'CH1', '600')     # set CH0 to 600 mV and CH1 to 600 mV

This function queries or sets the amplitude of the specified channels in mV. If there is only channel argument the function will return the amplitude of the specified channel. If there are two or four arguments the specified amplitude in mV will be set for specified channel.

Allowed channels: 'CH0', 'CH1'

Range (Spectrum M4I 6631 X8): 80 mV2500 mV

Range (Insys FM214x3GDA): 80 mV260 mV


awg_visualize()

awg_visualize()    # visualize the AWG card buffer (2D plot)

This function visualizes the AWG card buffer and can be called only without arguments. For the 'Single', 'Multi', and 'Single Joined' card mode, the complete buffer will be shown. For the 'Sequence' card mode only first two points will be shown. It is not recommended to use this function for very low repetition rate (less than 10 kHz) of awg_pulse_sequence() in the 'Sequence' card mode.


awg_pulse_list()

awg_pulse_list()    # -> list of str; pulse sequence

This function can be called only without arguments and it returns the declared pulse sequence as an array.


awg_clear()

awg_clear()

This is a special function for clearing pulse array {self.pulse_array} and other status flags of the device module. The function is usually used in GUI applications that use the device module.


awg_clear_pulses()

awg_clear_pulses()

This is a special function for clearing pulse array {self.pulse_array} and other status flags of the device module when the card is opened. The function is usually used in GUI applications that use the device module.


awg_correction(**kargs)

# measured resonator profile, applied only to the high-amplitude (pi) pulses
awg_correction(only_pi_half = 'True', coef_array = [bl, a1, x1, w1, a2, x2, w2, a3, x3, w3], \
               low_level = 16, limit = 23)

# ideal RLC resonator, amplitude + phase, applied to every swept pulse
awg_correction(only_pi_half = 'False', model = 'ideal', f0 = 9700, q_factor = 88, \
               phase_correction = 'True', low_level = 16, limit = 23)

# measured complex transfer (magnitude AND phase), applied to every swept pulse
awg_correction(only_pi_half = 'False', model = 'measured', f0 = 9700, \
               meas_freq = freq_MHz, meas_H = H_complex, low_level = 16, limit = 23)

This function (Insys FM214x3GDA and Spectrum M4I 6631 X8) enables resonator-profile predistortion of the frequency-swept pulses ('WURST' and 'SECH/TANH', see awg_pulse()) so that the effective excitation B1 is equalized across the swept band. Each sample of a swept pulse is mapped to its instantaneous chirp frequency (flipped high-frequency-first for an LO − RF bridge) and its amplitude — and optionally phase — is corrected accordingly. The correction has no effect on non-swept pulses.

There are two correction models:

  • model = 'measured' (default) — the amplitude is corrected with the measured resonator magnitude profile, given as a triple-Lorentzian fit in coef_array = [bl, a1, x1, w1, a2, x2, w2, a3, x3, w3] (the same coefficients written to correction.param by the resonator-tuning tool); no phase correction is applied. If, instead, a measured complex transfer function is supplied via meas_freq (absolute microwave frequency in MHz of each point) and meas_H (the complex transfer, e.g. a VNA S21 sweep), both the magnitude (1/|H|) and the phase (+angle(H), the LO − RF sign) are corrected: the real, non-Lorentzian magnitude ripple and the dispersion are equalized. Magnitude and unwrapped phase are interpolated per sample onto the chirp frequency; out-of-band samples clamp to the nearest measured point.
  • model = 'ideal' — the amplitude (1/|H|) and, when phase_correction = 'True', the phase are corrected with an ideal RLC resonator transfer function H = 1 / (1 + i Q (ν/f0 − f0/ν)), where f0 is the resonator centre frequency in MHz and q_factor the loaded Q. The phase term is +angle(H), the sign for an LO − RF (lower-sideband) bridge; verify it once on the bench for your setup.

The minimum effective B1 is clamped by the ratio low_level / limit (both in MHz), which also caps the amplitude boost applied to the band edges. only_pi_half = 'True' restricts the correction to the high-amplitude (pi) pulses (those with amplitude coefficient > 1 in awg_pulse()); only_pi_half = 'False' applies it to every swept pulse.

Note

The amplitude clamp equalizes the band only when the sweep is centred on the resonator (i.e. the microwave sweep is symmetric about f0). A sweep that sits entirely off resonance is clamped to a flat correction.

Allowed model: 'measured', 'ideal'

Default (Insys FM214x3GDA, Spectrum M4I 6631 X8): f0 = 9700 MHz, q_factor = 88, low_level = 16, limit = 23


awg_correction_off()

awg_correction_off()    # disable resonator-profile correction

This function disables the resonator-profile correction set by awg_correction(); subsequent swept pulses are sent as programmed. The function is usually used in GUI applications that use the device module.


awg_test_flag(flag)

awg_test_flag('test')    # change test mode

This is a special function for changing test mode. The function is usually used in GUI applications that use the device module.

Allowed: 'None', 'test'